Skip to content

Public FDA record

Z-1631-2014: ECATS E 850 Bariatric Bed, True Air Technologies, Inc.

Recall record Z-1631-2014 for SCM True Air Technologies LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1631-2014
Company in FDA record
SCM True Air Technologies LLC
FDA classification date
May 19, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The firm is recalling the products due to lack of design controls, lack of a Device Master Record, and failure to approve specific vendors.

Source facts

Action
SCM True Air Technologies LLC sent an Urgent Medical Device Recall letter to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to remove the affected product from use and return to True Air Technologies immediately. The units should be decontaminated and bagged with a certification of disinfection before return to True Air Technologies. True Air will provide replacement units to replace the returned units. Customers were instructed to refer to the Medical Device Recall Return Response Form attached for acknowledgement of receipt and contact information for this voluntary recall. Customers with questions should call 800-682-7163.
Initiated
2014-03-18
Root cause
Nonconforming Material/Component
Terminated
2018-06-11
Distribution
Nationwide Distribution including GA, KY, and MI.
Recall number
Z-1631-2014
Recall status
Terminated
CDRH record ID
126957
FDA classified
2014-05-19
Recall event number
67924

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗