At doses of less than 100 mJ, in custom and auto repeat modes, a software error in the excimer laser phototherapy system causes the device to fail to cease firing as long as the foot pedal or handpiece switch is depressed. This malfunction can cause overtreatment, which could result in strong erythema, and possibly cause blisters.
Source facts
Action
On 08/23/2019, Medical Device Correction was mailed to customers. Customers were informed that the recalling firm's service department would contact them to replace the software. Unitl the software is replaced, do not use doses less than 100 mJ. During service visits customers were informed that software was updated to 1.06 ZRPL
Initiated
2019-08-23
Root cause
Software change control
Terminated
2021-05-05
Distribution
U.S. Nationwide distribution in the states of CA, FL, OR, MS, NY.
Recall number
Z-1631-2020
Recall status
Terminated
CDRH record ID
180264
FDA classified
2020-04-01
Recall event number
85200