Skip to content

Public FDA record

Z-1632-2010: Leica Footswitch, Type B used in conjunction with Leica Surgical Microscopes - Models M620, M844/820, M52

Recall record Z-1632-2010 for Leica Microsystems, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1632-2010
Company in FDA record
Leica Microsystems, Inc.
FDA classification date
May 20, 2010
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The pedals of the Footswitch Type B stick and the Zoom/Focus moves to the end position.

Source facts

Action
Leica Microsystems sent Medical Device Correction letters dated 3/29/10 to the affected customers via first class mail on the same date, informing them of the potential for injury to the patient's eye if the cross pedal of the footswitch sticks. They were informed that their local Leica Microsystems representative will contact them to arrange for the replacement of the cross pedal. The letters also provided the interim necessary steps required to avoid the risk of patient injury while they are awaiting the cross pedal replacement, i.e. do not use the affected footswitches with a surgical microscope that has a BIOM accessory. The accounts were requested to complete the attached Acknowledgement Form, listing the serial numbers of the units at their facility and fax it back to Leica at 1-847-607-3147. A company contact was provided in the letter for questions.
Initiated
2010-03-29
Root cause
Other
Terminated
2011-02-15
Distribution
Worldwide distribution -- U.S. (nationwide), Canada, Australia, Europe and Asia
Recall number
Z-1632-2010
Recall status
Terminated
CDRH record ID
90391
FDA classified
2010-05-20
Recall event number
55241

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗