Reason stated in the source
A performance issue with the chest depth needle valve (CDNV) may arise where turning the knob will not deliver any compression depth to the piston during the first 5 revolutions of the knob.The last to revolution of the knob could deliver a full 8cm of compression.
Source facts
- Action
- The customer was notified via email on 2/13/18. The customer has been instructed to identify the location of the affected units (if distributed, make arrangements to have unit returned for service), perform the CDNV replacement (instructions will be provided), and test the device to the Final Test Procedure (TS190). The customer should also record on the provided spreadsheet the SN of the CDNV replaced in the CCV device, date of replacement. A column is provided to check-off indicating the device was tested and accepted to the Final Test Procedure (TS190). Return the completed spreadsheet and the CDNVs that were replaced for evaluation.
- Initiated
- 2018-02-13
- Root cause
- Nonconforming Material/Component
- Terminated
- 2019-04-03
- Distribution
- Distributed to China.
- Recall number
- Z-1632-2018
- Recall status
- Terminated