Open original FDA record ↗Recall record
- FDA identifier
- Z-1633-2016
- Company in FDA record
- Siemens Healthcare Diagnostics, Inc.
- FDA record creation date
- May 10, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Siemens Healthcare Diagnostics has received customer complaints regarding QC shifts and imprecision after one day when using Dimension SIRO lot EB6064. They have confirmed SIRO lot # EB6064 does not meet the open well stability claim of 2 days.
Source facts
- Action
- Siemens sent an "URGENT MEDICAL DEVICE RECALL" letter dated February 2016, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Actions to be Taken by the Customer: Discontinue use of and discard the kit lot listed in Table 1. Please review this letter with your Medical Director. Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. If you have received any complaints of illness or adverse events associated with the product listed in Table 1, immediately contact your local Siemens Customer Care Center or your local Siemens technical support representative. Please retain this letter with your laboratory records, and forward this letter to those who may have received this product. We apologize for the inconvenience this situation may cause. If you have any questions, please contact your Siemens Customer Care Center or your local Siemens technical support representative. For further questions, please call (302) 631-0528.
- Initiated
- 2016-02-09
- Root cause
- Under Investigation by firm
- Terminated
- 2018-06-12
- Distribution
- Worldwide distribution and US including Puerto Rico and to the states of : TX, FL, CA, LA, DC, GA, IL, AR, WI, MI and MD.
- Recall number
- Z-1633-2016
- Recall status
- Terminated
- CDRH record ID
- 144211
- FDA record created
- 2016-05-10
- Recall event number
- 73442