Open original FDA record ↗Recall record
- FDA identifier
- Z-1633-2020
- Company in FDA record
- Physio-Control, Inc.
- FDA classification date
- Apr 1, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Automated External Defibrillators may not detect a patient connection due to mechanical wear-through of the contact plating on the therapy connector. The wear-through could result in the device not recognizing a patient is connected which could result in a delay in treatment
Source facts
- Action
- The firm is planning on providing consignees on March 20, 2020 with an "Urgent Medical Device Safety Notice & Correction" letter concerning the LIFEPAK 500 AED. In addition to informing consignees about the recalled device, the recall notifications requested consignees take the following action: If you experience a "CONNECT ELECTRODES" voice prompt with the LIFEPAK 500, immediately remove and reinstall the electrodes to the device or replace the electrodes with your spare electrodes and check patient connection. If CONNECT ELECTRODES" voice prompt continues, immediately obtain a backup device and remove the LIFEPAK 500 from use. " Continue to perform the External Test Load Test per Maintenance section of the LIFEPAK 500 Operating Instructions and your local protocols as this test may identify this issue prior to being used on a patient. If during this test, a CONNECT ELECTRODES voice prompt message is received, immediately remove the device from service. " For any devices that are still within their expected life (meaning it has not been more than eight years since the original purchase of the new device from Physio-Control/Stryker or its authorized distributors), include the attached additional warnings and cautions as supplemental labeling for your device. " Review your attached affected device list and confirm if your device is still in service, transferred to a new location or is no longer in service. Please return the confirmation sheet by fax at 1-866-448-9567. If you have questions regarding the continued safe use of your products, please contact our Technical Support team at Stryker at 1 800 787 9537 and select option 2, 8:00 A.M. to 6:00 P.M. (Eastern time), Monday Friday. More information from FDA regarding Legacy AED Systems can be accessed at https://www.fda.gov/medical-devices/cardiovascular-devices/automated-externaldefibrillators-aeds. If you have any questions about this matter contact Stryker at 1 800 787 9537, option
- Initiated
- 2020-03-20
- Root cause
- Device Design
- Terminated
- 2024-01-25
- Distribution
- Worldwide distribution - US Nationwide distribution and countries of Afghanistan, American Samoa, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Barbados, Belarus, Belgium, Bermuda, Bolivia, Bosnia and Herzegovina, Brazil, Canada, Cayman Islands, Chile, China, Columbia, Congo, Cook Islands, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Finland, France, French Guiana, French Polynesia, Georgia, Germany, Ghana, Greece, Guam, Guatemala, Haiti, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Islamic Republic of, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kenya, Korea, Korea, Republic of Kosovo, Kuwait, Kyrgyzstan, Latvia, Lebanon, Lithuania, Macedonia, the former Yugoslav Republic of, Madagascar, Malaysia, Mexico, Netherlands, New Caledonia, New Zealand, Nicaragua, Niger, Nigeria, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, R¿union, Romania, Russian Federation, San Marino, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Province of China, Togo, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, Bolivarian Republic of, Vietnam, Virgin Islands, British, Virgin Islands, U.S., Zimbabwe.
- Recall number
- Z-1633-2020
- Recall status
- Terminated
- CDRH record ID
- 180255
- FDA classified
- 2020-04-01
- Recall event number
- 85194