There is a potential for inaccuracies in blood gas readings/measurements. There is a potential risk of respiratory alkalosis or cerebral vasoconstriction with low paCO2. If the Hemoglobin and Hematocrit measurements are not accurate, there is a potential risk of unnecesary blood transfusion with resulting risk of infection or inflammatory reaction. BPM shipped without meeting manufacturing accep
Source facts
Action
On 5/9/14, TCVS placed a phone call to the one affected customer regarding the recall. The phone call explained the reason for the recall, identification of affected product, potential hazard, and instructions for return of affected product.
Initiated
2014-05-09
Root cause
Release of Material/Component prior to receiving test results
Terminated
2014-07-11
Distribution
one customer in IA
Recall number
Z-1634-2014
Recall status
Terminated
CDRH record ID
127202
FDA classified
2014-05-19
Recall event number
68240