Open original FDA record ↗Recall record
- FDA identifier
- Z-1635-2015
- Company in FDA record
- Stryker Howmedica Osteonics Corp.
- FDA classification date
- May 19, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The peel strength of the inner blister may have been below internal validated requirements. The strength of the seal is directly related to long term shelf life of the sterile barrier.
Source facts
- Action
- The firm, Stryker Orthopaedics, notified their Branches/Agencies via email on 8/15/2014 and sent an "Urgent Product Recall" letter dated 8/15/2014 with a Product Recall Acknowledgement Forms to their Branches/Agencies and consignees. The letter describes the product, issue, potential hazards and actions to be taken. The consignees were instructed to complete and return the attached Product Recall Acknowledgment Form within 5 days via fax to 855-251-3635; and return the affected product to the attention of Regulatory Compliance, Stryker Orthopaedics, 325 Corporate Drive, Mahwah NJ 07430. If you have any questions, feel free to contact Sr. Regulatory Compliance Specialist to 201-831-5826.
- Initiated
- 2014-08-15
- Root cause
- Under Investigation by firm
- Terminated
- 2018-01-05
- Distribution
- Worldwide Distribution: US (nationwide) including PR and countries of: Australia, Canada, China, Brazil, France, Hong Kong, India, Italy, Mexico, Netherlands, Singapore, South Africa, Sweden, and Spain.
- Recall number
- Z-1635-2015
- Recall status
- Terminated
- CDRH record ID
- 135950
- FDA classified
- 2015-05-19
- Recall event number
- 70994