Skip to content

Public FDA record

Z-1635-2026: MiniMed" 780G - MMT-1884 MiniMed" 780G - MMT-1886 With Software 6.60 and 6.61

Recall record Z-1635-2026 for Medtronic MiniMed, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1635-2026
Company in FDA record
Medtronic MiniMed, Inc.
FDA classification date
Mar 24, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Three software defects (Pump Error 53, BG check, and Critical Pump Error (Open Book Image)) that occurred as a result of software updates (versions 6.60, 6.61, and 6.62), which could result in a risk of hypoglycemia or hyperglycemia due to inaccuracies in insulin delivery (either under-delivery, over-delivery, or cessation of delivery).

Source facts

Action
On 01/30/2026, the firm emailed an initial "URGENT MEDICAL DEVICE CORRECTION" Letter to customer informing them of two issues affecting the MiniMed 780 Insulin Pump Model MMT-1884 that contain software version 6.60 and 6.61. The issues are: 1-Pump Error 53 which indicates that insulin delivery is being stopped. 2-Blood Glucose (BG) Check functionality may not conduct verification/performance checks. Due to a software anomaly, this verification may not occur as expected when the SG reading is above 400 mg/dL or below 50 mg/dL. On 02/20/2026, an "UPDATED: URGENT MEDICAL DEVICE CORRECTION" letter via email informing customer of a third issue discovered with the implementation of software version 6.62, Critical Pump Error (Open Book Image) which may stop insulin delivery. Customers are instructed to: If using Pump Software Ver 6.60 or 6.61: 1. Ensure MiniMed" Mobile app is updated to version 3.1.1 (iOS) and 3.1.2 (Android) or higher. If it is not, complete that update. Click Here for instructions on updating your app. Note: To identify your current Mobile App version, Go to Menu (top left), tap on About . 2. Update MiniMed" 780G insulin pump software to version 6.62. Click Here to refer to the detailed software update guide and update your pump software. Note: For instructions on how to find your pump software version, Click Here. 3. If customer is wearing the Instinct sensor during the update to software version 6.62, immediately restart the Instinct sensor. If customer has already updated to Pump Software Ver 6.62: 1. if wearing an instinct sensor, unpair and re-pair sensor for uninterrupted therapy For assistance with updating MiniMed" Mobile app to the latest version or your insulin pump software to version 6.62, or or experiencing Critical Pump Error (Open Book Image), any questions about the process, please contact 24-Hour Technical Support Team at 1-800-646-4633, option 1.
Initiated
2026-01-30
Root cause
Software change control
Distribution
Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH,NJ, NM, NV, NY, OH,OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY and the country of Israel.
Recall number
Z-1635-2026
Recall status
Open, Classified
CDRH record ID
218770
FDA classified
2026-03-24
Recall event number
98444

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗