Open original FDA record ↗Recall record
- FDA identifier
- Z-1636-2014
- Company in FDA record
- Smith & Nephew Inc
- FDA classification date
- May 20, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
A manufacturing error resulted in the set screw being inserted too deep in the affected nails.
Source facts
- Action
- The firm initiated the recall via an "Urgent - Product Recall 1st Notification - Urgent" letter and e-mail on 04/04/2014, identifying the product, explaining the problem, and the potential risk associated with using the product. Customers are to inspect their inventory, and locate and quarantine any affected product. Additionally, the Inventory Return Certification form should be completed and returned along with any affected product identified, once a Return Authorization number has been obtained from Smith & Nephew's Safety Affairs Department.
- Initiated
- 2014-04-04
- Root cause
- Process control
- Terminated
- 2017-07-01
- Distribution
- Worldwide Distribution -- USA, France, Germany, Sweden, Switzerland, UK, Canada, Australia, Japan, China, India, Dubai, South Africa, South Korea, Mexico, and Venezuela.
- Recall number
- Z-1636-2014
- Recall status
- Terminated
- CDRH record ID
- 126851
- FDA classified
- 2014-05-20
- Recall event number
- 68043