Z-1636-2026: Tandem pumps are battery-operated infusion pumps capable of both basal and bolus delivery of insulin. The

Recall record Z-1636-2026 for Tandem Diabetes Care, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1636-2026
Company in FDA record
Tandem Diabetes Care, Inc.
FDA classification date
Mar 25, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The Czech language user guide contained multiple translation errors. The most significant error was that it incorrectly instructs users to verify that the infusion set is connected to the body, when it should instead direct users to confirm that the infusion set is not connected to the body. Following this instruction could result in unintended insulin delivery, potentially leading to a hypoglycemic event.

Source facts

Action
On January 30, 2026, Tandem Diabetes Care issued a Urgent Medical Device Recall Notice to affected consignees via E-Mail. Tandem ask consignees to take the following actions: 1. Discard your existing User Guide. 2. Refer to the corrected User Guide version 7.8.1 which is available online at www.tandemdiabetes.com/cs-cz/support/resources/documents for accurate instructions on using your t:slim X2 Insulin Pump. 3. Physical copies of the corrected User Guide are being produced to replace the incorrect versions. If you would like a physical copy of the corrected User Guide, you may contact A. Import s customer service line to arrange for a replacement. 4.As requested above, please complete the online form acknowledging receipt of this notice. 5. Should you experience any device-related serious incidents, please report them to your distributor, a local contact point, or the State Institute for Drug Control.
Initiated
2026-01-30
Root cause
Process change control
Distribution
International distribution in the countries of Czech Republic, Slovakia.
Recall number
Z-1636-2026
Recall status
Open, Classified
CDRH record ID
218867
FDA classified
2026-03-25
Recall event number
98478

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗