Open original FDA record ↗Recall record
- FDA identifier
- Z-1637-2016
- Company in FDA record
- Bradshaw Medical Inc.
- FDA classification date
- May 10, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
A component failure could cause a situation where the applied torque was significantly higher than the calibrated torque setting.
Source facts
- Action
- Consignees were sent a Bradshaw Medical "Product Withdrawal Notice" letter, date marked starting February 2, 2015. The letter specified the Part Number, Lot Number (s) and Shipped date for each consignee. The letter described the product and advised consignees to discontinue use of the product. Requested consignees to complete and return the Customer /Distribution verification Form to BMI Regulatory Affairs FAX (262) 925-1380 and to further send the "Withdrawal Notice and Customer Verification Form" to all customers to whom they have distributed the product.
- Initiated
- 2015-02-02
- Root cause
- Device Design
- Terminated
- 2020-01-10
- Distribution
- US Nationwide Distribution in the states of: CA, FL, IN, MA, MI, OH, PA, TN UT, VA and the country of UNITED KINGDOM
- Recall number
- Z-1637-2016
- Recall status
- Terminated
- CDRH record ID
- 145265
- FDA classified
- 2016-05-10
- Recall event number
- 73873