Z-1637-2026: Blue Ventilator Adapter Module, single-patient use circuit, Product Code M07937. Accessory to Baxter Vol

Recall record Z-1637-2026 for Baxter Healthcare Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1637-2026
Company in FDA record
Baxter Healthcare Corporation
FDA classification date
Apr 1, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Urgent Medical Device Correction for the Volara system single-patient use circuit due to the potential that patients or caregivers may be unaware of a possible decrease in oxygen levels (oxygen desaturation), or potential lung tissue injury from overexpansion (barotrauma) when the Volara device is used in the home care environment. This correction is providing the home caregiver specific instructions for use.

Source facts

Action
Baxter issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 2/18/2026 via email. The notice explained the issue, hazard involved, and requested the following: "Actions to be Taken by Customers 1. Continue your Volara therapy as prescribed by your doctor. 2. Refer to the enclosed IFU addendum for additional information related to use in the home care environment. The following warning was added for consideration related to this issue: Warning - Patients that may have difficulty clearing secretions from the upper airway may require specialized therapy (for example, separate suction unit) in addition to Volara therapy. Please consult your physician to determine if additional therapy is appropriate. 3. Monitor for signs of respiratory distress (increased breathing rate; wheezing; bluish color around the mouth, on the lips, or on the fingernails; changes in alertness; or drop in oxygen saturation level, etc.) during Volara therapy when used in-line with your ventilator. If signs of respiratory distress occur, stop the Volara therapy immediately. If you experience signs of respiratory distress and do not see improvement, please seek medical attention. 4. If you have questions or would like assistance with your Volara therapy in-line with a ventilator, please contact the Baxter Advanced Respiratory Clinical Support team at 800-397-9071, available 24 hours a day, 7 days a week. Clinical support can be provided via the phone, or Baxter can arrange for an in-person visit with a clinical patient trainer. 5. As part of the additional training, you will be asked to sign your confirmation of understanding of the training material and content of the IFU addendum." For general questions regarding this communication, please contact Baxter Advanced Respiratory Clinical Support team at 800-397-9071, available 24 hours a day, 7 days a week.
Initiated
2026-02-18
Root cause
Process control
Distribution
Florida
Recall number
Z-1637-2026
Recall status
Open, Classified
CDRH record ID
218810
FDA classified
2026-04-01
Recall event number
98458

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗