Laser aimer assemblies were shipped without certification of the overall assembly or submitting a laser product report and therefore do not comply with the US regulations for laser products.
Source facts
Action
GE Healthcare Surgery Planned Action(s) to repair the defect or to bring the product into compliance:
The Center for Devices and Radiological Health (CDRH) understands that the CAP involving 9 units consisted of the following:
1. Replacing the tube side cover with the laser aimer assembly with a tube
side cover without a laser aimer assembly.
This correction brought the systems into compliance with applicable US regulations. For further questions please call (801) 536-4615.
Initiated
2014-05-22
Root cause
Radiation Control for Health and Safety Act
Terminated
2015-06-24
Distribution
US Distribution to the states of: MD, NC, MT, NJ, UT and NY
Recall number
Z-1638-2015
Recall status
Terminated
CDRH record ID
137256
FDA classified
2015-06-08
Recall event number
71298