Z-1639-2012: BD Focal Point"230V / 11 FOV under 5%, Catalog number: 490648. The BD Focal Point" Slide Profiler (fo

Recall record Z-1639-2012 for Tripath Imaging, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1639-2012
Company in FDA record
Tripath Imaging, Inc.
FDA classification date
May 21, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The cable tie mounts that adhere to the inside of the FocalPoint do not adhere as intended and can therefore become detached.

Source facts

Action
BD Diagnostics first notified consignees by letter on October 6, 2011. BD Diagnostics also sent a letter dated February 2012, to all affected consignees. The letter identified the product, the problem, and the action to be taken by the consignee. Field Representatives were advised to inspect all cables, adhesive cable ties and connections to ensure they are secure. Inspections and repairs, if necessary, would be carried out at the next service call for each instrument, or no later than the next regularly scheduled preventative maintenance visit. For questions consignees were instructed to contact BD Diagnostics Technical Support at 1-877-822-7771 option 2 in the US, their local distributor, or their local BD office for any questions. For questions regarding this recall call 918-563-1624.
Initiated
2011-10-06
Root cause
Component change control
Terminated
2012-12-03
Distribution
Worldwide Distribution - USA including AZ, CT, IL, IN, FL, MN, NC, OK, OR, and TX and the countries of Australia, Belgium, Brazil, Canada, Japan and Thailand
Recall number
Z-1639-2012
Recall status
Terminated
CDRH record ID
108122
FDA classified
2012-05-21
Recall event number
61444

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗