Skip to content

Public FDA record

Z-1639-2022: AggreGuide A-100 Instrument

Recall record Z-1639-2022 for Aggredyne, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1639-2022
Company in FDA record
Aggredyne, Inc.
FDA record creation date
Aug 26, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The firm identified a software design issue where the improper shut down of the instrument resulted in speeds.csv file contents of the software version 5.1 being erased. Due to this issue, the target RPM for the assay is 0 and the motor output required to achieve this RPM is also 0. Therefore, when the assay is run, the software applies 0 Volts to the motor and performs a 0 RPM assay. Because of this, there is no mixing of the blood with agonist and no moving of aggregates happens in front of the optical detectors. The resulting PAI is 0 and the result is displayed as Low.

Source facts

Action
On 7/23/2019, a Service Bulletin was sent to distributors and users informing them to follow the User Manual, which states: 1) To shut down the instrument: Press the Shut Down button and wait. 2) Continue to wait for the screen message stating that it is safe to turn off your computer (this may take 15 seconds or more). 3) Turn off power switch only after the message "It is now safe to turn off your computer" is displayed. 4) A future communication from the recalling firm will be provided when the software update is available. On 7/8/22 updated Service Bulletin was sent to the distributor notifying them the software update was available. Field technicians will upgrade affected devices with software version 5.2.1.1.
Initiated
2019-07-23
Root cause
Software design
Distribution
International distribution to the country of South Korea.
Recall number
Z-1639-2022
Recall status
Completed
CDRH record ID
194963
FDA record created
2022-08-26
Recall event number
90588

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.