Open original FDA record ↗Recall record
- FDA identifier
- Z-1640-2008
- Company in FDA record
- Baxter Healthcare Corp.
- FDA classification date
- Aug 24, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Software anomaly results in inaccurate information being printed out on the Delivery Report, even though the compounder performed the compounding correctly.
Source facts
- Action
- Baxter sent the 3/28/08 Urgent Device Correction letter to affected customers, to the attention of the Director of Pharmacy, via first class mail to advise them of the potential inaccuracy of the Delivery Report due to a software anomaly. The accounts were instructed to review each delivery report for accuracy. If the report contains zero actual volumes the user must manually reprint the report. The letter outlined the process for reprinting the Delivery Report. Any questions were directed to the Baxter Nutritional Professional Services at 1-800-422-2751. The accounts were requested complete the enclosed reply form acknowledging receipt and understanding of the letter and dissemination of the information to their staff. A software upgrade will be provided. If you have questions about this letter or you did not receive this letter contact Baxter Nutrition Professional Services at 1-800-422-2751.
- Initiated
- 2008-03-28
- Root cause
- Software design
- Terminated
- 2008-08-25
- Distribution
- Arizona, Minnesota, Wisconsin, Puerto Rico and Canada
- Recall number
- Z-1640-2008
- Recall status
- Terminated
- CDRH record ID
- 69504
- FDA classified
- 2008-08-24
- Recall event number
- 47441