- FDA identifier
- Z-1640-2026
- Company in FDA record
- Olympus Corporation of the Americas
- FDA classification date
- Mar 25, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.
Source facts
- Action
- On February 25, 2026 URGENT: MEDICAL DEVICE REMOVAL letters were sent to customers. Actions to be taken: 1. Examine inventory and quarantine any affected devices 2. Cease usage of the product with immediate effect 3. Please contact Customer Service at 1-800-848-9024, option 2, to obtain a Return Material Authorization 4. Even if you no longer have the device, Olympus requests that you acknowledge receipt of this letter through the Olymps portal 5. Please forward this notice to other users who may have the affected products if you have further distributed it.
- Initiated
- 2026-02-25
- Root cause
- Nonconforming Material/Component
- Distribution
- Worldwide - US Nationwide and the countries of Canada, Mexico, EU, Australia, China, Singapore, Korea, Japan.
- Recall number
- Z-1640-2026
- Recall status
- Open, Classified
- CDRH record ID
- 218800
- FDA classified
- 2026-03-25
- Recall event number
- 98454