Z-1641-2008: CADD-MS"3 Ambulatory Infusion Pumps, Model 7400, Rx only, CE 0473, SN 499000. The device is designed fo

Recall record Z-1641-2008 for Smiths Medical MD, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1641-2008
Company in FDA record
Smiths Medical MD, Inc.
FDA classification date
Aug 24, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Smiths Medical has received reports that the device powers down without an alarm. Investigation concluded that if the battery cap is not fully tightened, the pump may power down and a brief "chirp" will sound. This may occur if the user has not sufficiently tightened the battery cap or if the battery cap is damaged. If a patient is asleep or does not regularly check their Pump, there is a potentia

Source facts

Action
A "Smith Medical Urgent Medical Device Correction" letter dated February 22, 2008 was distributed to consignees. The letter addressed to Risk/Safety Managers, Distributors, Clinicians, Patients and other Customers, instructs them to fully tighten the battery cap and to routinely monitor the pump for damage that may result in power down situations. Consignees are to return the "Urgent Medical Device Correction" form. If you have questions about the letter or did not received the letter contact Mike Herbert at 651-628-7049 or Jeff Esselman at 651-628-7920.
Initiated
2008-02-22
Root cause
Labeling design
Terminated
2011-01-11
Distribution
AZ CA FL GA HI IL KY MD MI MN NC NE NJ NM NY PA TN TX UT CANADA CHILE CZECH REPUBLIC GERMANY GREAT BRITAIN ITALY JAPAN NETHERLANDS NORWAY SWEDEN
Recall number
Z-1641-2008
Recall status
Terminated
CDRH record ID
69545
FDA classified
2008-08-24
Recall event number
47471

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗