Open original FDA record ↗Recall record
- FDA identifier
- Z-1641-2008
- Company in FDA record
- Smiths Medical MD, Inc.
- FDA classification date
- Aug 24, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Smiths Medical has received reports that the device powers down without an alarm. Investigation concluded that if the battery cap is not fully tightened, the pump may power down and a brief "chirp" will sound. This may occur if the user has not sufficiently tightened the battery cap or if the battery cap is damaged. If a patient is asleep or does not regularly check their Pump, there is a potentia
Source facts
- Action
- A "Smith Medical Urgent Medical Device Correction" letter dated February 22, 2008 was distributed to consignees. The letter addressed to Risk/Safety Managers, Distributors, Clinicians, Patients and other Customers, instructs them to fully tighten the battery cap and to routinely monitor the pump for damage that may result in power down situations. Consignees are to return the "Urgent Medical Device Correction" form. If you have questions about the letter or did not received the letter contact Mike Herbert at 651-628-7049 or Jeff Esselman at 651-628-7920.
- Initiated
- 2008-02-22
- Root cause
- Labeling design
- Terminated
- 2011-01-11
- Distribution
- AZ CA FL GA HI IL KY MD MI MN NC NE NJ NM NY PA TN TX UT CANADA CHILE CZECH REPUBLIC GERMANY GREAT BRITAIN ITALY JAPAN NETHERLANDS NORWAY SWEDEN
- Recall number
- Z-1641-2008
- Recall status
- Terminated
- CDRH record ID
- 69545
- FDA classified
- 2008-08-24
- Recall event number
- 47471