Skip to content

Public FDA record

Z-1641-2014: Portex Low Dead Space Connector with Sideport, 3.5mm Product Usage: The connector is used to attach t

Recall record Z-1641-2014 for Smiths Medical ASD, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1641-2014
Company in FDA record
Smiths Medical ASD, Inc.
FDA classification date
Jun 11, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

One lot of 3.0mm sized Connectors, Product Reorder No. 10035-05PS, Lot No. 2553426 were placed in packages labeled as size 3.5mm.

Source facts

Action
Smiths Medical sent an Urgent Medical Device Recall Notice dated April 10, 2014, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to inspect their inventory and quarantine all affected products and to complete the attached Response Form and return it to Smith Medical within 7 days of receipt of the notice. Customers with questions were instructed to contact Smiths Medical's Customer Service Department at +1-800-258-5361.
Initiated
2014-04-10
Root cause
Packaging process control
Terminated
2016-11-29
Distribution
US distribution in the state of North Carolina.
Recall number
Z-1641-2014
Recall status
Terminated
CDRH record ID
126747
FDA classified
2014-06-11
Recall event number
67991

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗