Open original FDA record ↗Recall record
- FDA identifier
- Z-1641-2018
- Company in FDA record
- Medtronic Sofamor Danek USA Inc
- FDA record creation date
- May 4, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The firm received complaints of drill bits breaking during use with the Prestige LP(TM) Cervical Disc System. Subsequent investigation demonstrated the interaction of the drill guide and the drill bit may lead to drill bit breakage.
Source facts
- Action
- Medtronic distributed a voluntary removal notice via FedEx courier service to impacted U.S. consignees on 13MARCH 2018. The notice requested the following: "Instructions to Customers: 1.) Please locate and remove the impacted product from normal storage locations. Do not use this product. 2.) Complete the enclosed Customer Confirmation Form and return via fax to 763-367-8134, or via e-mail to rs.sdmriskmgt@medtronic.com. 3.) As applicable, contact your Medtronic Sales Representative to coordinate the return of affected product."
- Initiated
- 2018-03-13
- Root cause
- Under Investigation by firm
- Terminated
- 2020-11-02
- Distribution
- US Nationwide Distribution
- Recall number
- Z-1641-2018
- Recall status
- Terminated
- CDRH record ID
- 162930
- FDA record created
- 2018-05-04
- Recall event number
- 79667