Open original FDA record ↗Recall record
- FDA identifier
- Z-1642-2009
- Company in FDA record
- Stryker Howmedica Osteonics Corp.
- FDA classification date
- May 21, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Packaging burst test values for the Restoration Modular Cone Body family were found to be incorrectly transferred from the finalized validation to the work specification. Testing showed that burst test results for recalled products fell below the validated parameters.
Source facts
- Action
- Urgent Product Recall letters were sent via Federal Express on April 2, 2009 to Stryker branches, Chief of Orthopaedics, Hospital Rick Management and Surgeons. The letter stated the issue, the potential hazards, and the risk mitigation. Customers were asked to examine their inventory and hospital locations to identify the affected products; retrieve all effected products and return it to their branch or agency warehouse for reconciliation; reconcile all products on the attached Product Accountability Form and fax a copy to the number listed; return the affected product via the Inter-Org Transfer and ship it to the attention of the individual listed; attach the fluorescent orange PRODUCT REMEDIATION sticker to their return, indicating the Product Remediation #, and mark the box with the words "PRODUCT RECALL". Questions should be directed to Colleen O'Meara at 201-831-5970.
- Initiated
- 2009-04-02
- Root cause
- Process control
- Terminated
- 2012-08-27
- Distribution
- Nationwide Distribution
- Recall number
- Z-1642-2009
- Recall status
- Terminated
- CDRH record ID
- 81477
- FDA classified
- 2009-05-21
- Recall event number
- 51868