Open original FDA record ↗Recall record
- FDA identifier
- Z-1642-2014
- Company in FDA record
- Fujifilm Medical System USA, Inc.
- FDA classification date
- May 21, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The MXR-35 X-ray Generator's labeling did not include the certification statement as required by 21 CFR Part 10 10.2.
Source facts
- Action
- Fujifilm will replace the violative labels with replacement labels provided by DRGEM, to bring the X-ray generator into compliance. You stated that you expect for all nine installed units to be corrected by the end of May 2014. The seven units in the Stamford, CT warehouse will be corrected when the generators are delivered to customer sites. Under 21 CFR 1003.31(c)(2), this exemption is granted subject to the following conditions: 1. You must correct the missing certification statement for all units shipped into the US within the timeline you specified, by the end of May 2014. 2. You must ensure that all future units that will be delivered and installed in the US will be in compliance with the labeling requirement, per 21 CFR 1010.2.
- Initiated
- 2012-06-29
- Root cause
- Radiation Control for Health and Safety Act
- Terminated
- 2018-06-01
- Distribution
- Worldwide Distribution - US Distribution including the states of TX and KY and Internationally to China.
- Recall number
- Z-1642-2014
- Recall status
- Terminated
- CDRH record ID
- 126664
- FDA classified
- 2014-05-21
- Recall event number
- 67958