Reason stated in the source
The firm discovered a problem where the selected patient images on the device may display an additional image from another patient or study.
Source facts
- Action
- The recalling firm issued a Customer Safety Advisory via certified mail to affected customers per Update Instructions IM011/08/S. The letter informs customers of the potential issue and provides instructions to detect its occurrence. A software update to correct the issue is already available. A service representative will visit affected sites to install the software update. If you have questions concerning the advisory letter or you did not receive a letter contact Alexander Vogel at +49 (9131) 84-3840 or alexander.vogel@siemens,com and make reference to complaint 08-1KM-0073 date 2008-03-06.
- Initiated
- 2008-03-10
- Root cause
- Software design