Open original FDA record ↗Recall record
- FDA identifier
- Z-1643-2015
- Company in FDA record
- Philips Medical Systems, Inc.
- FDA classification date
- May 21, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
A problem (the dose computed in planning mode is calculated incorrectly ) has been detected in the Philips Pinnacle3 software version 10.0 that, if it were to re-occur, could pose a risk for patients or users. Specifically, the dose engine is being passed the wrong snout position. The snout position is used to determine the penumbra of the beam which includes the calculation of the source si
Source facts
- Action
- Philips Healthcare sent a "Field Safety Notification PROST-T-300-01-06" letter dated 20 April 2015 to their customers. The letter described the Problem Description, Hazard Involved, How to Identify Affected Products, Action to be Taken by Customer / User, Actions Planned by Philips and Further Information and Support (Contact Philips Customer Care Solutions Center at 1-800-722-9377).
- Initiated
- 2015-04-23
- Root cause
- Software design
- Terminated
- 2016-10-05
- Distribution
- US Nationwide Distribution in the states of MO, FL, OH
- Recall number
- Z-1643-2015
- Recall status
- Terminated
- CDRH record ID
- 136154
- FDA classified
- 2015-05-21
- Recall event number
- 71117