Open original FDA record ↗Recall record
- FDA identifier
- Z-1643-2018
- Company in FDA record
- Beckman Coulter Inc.
- FDA record creation date
- May 4, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Beckman Coulter has determined that there is a potential for incorrect settings to be installed on North American iChemVELOCITY Urine Chemistry Analyzers. 1. The installation of International settings can result in the generation of erroneous, false low results for some of the analytes at some concentrations. 2. The installation of colors different from those stated in the Instructions for Use (Colorless, Straw, Yellow, Amber, Red, Blue) for output settings will result in incorrect reporting of colors. For example, if Green is the color choice for the output setting instead of Blue, Green will be reported.
Source facts
- Action
- Beckman Coulter sent an Urgent Field Safety Notice letter dated February 22, 2018 to the customers. The letter identified the affected product, problem and actions to be taken. Beckman Coulter will schedule a site visit to verify the reporting units of measurement for the laboratory within 90 days of this recall initiation. For questions call 800-854-3633.
- Initiated
- 2018-03-02
- Root cause
- Under Investigation by firm
- Terminated
- 2023-07-27
- Distribution
- Worldwide Distribution - US Nationwide and the countries of Brazil, Canada, Malaysia, Singapore, Taiwan, Thailand, Turkey
- Recall number
- Z-1643-2018
- Recall status
- Terminated
- CDRH record ID
- 162677
- FDA record created
- 2018-05-04
- Recall event number
- 79575