Z-1643-2018: BECKMAN COULTER iChemVELOCITY, Urine Chemistry System, Catalog Numbers: 800-3530, 800-7167, 800-7719, 80

Recall record Z-1643-2018 for Beckman Coulter Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1643-2018
Company in FDA record
Beckman Coulter Inc.
FDA record creation date
May 4, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Beckman Coulter has determined that there is a potential for incorrect settings to be installed on North American iChemVELOCITY Urine Chemistry Analyzers. 1. The installation of International settings can result in the generation of erroneous, false low results for some of the analytes at some concentrations. 2. The installation of colors different from those stated in the Instructions for Use (Colorless, Straw, Yellow, Amber, Red, Blue) for output settings will result in incorrect reporting of colors. For example, if Green is the color choice for the output setting instead of Blue, Green will be reported.

Source facts

Action
Beckman Coulter sent an Urgent Field Safety Notice letter dated February 22, 2018 to the customers. The letter identified the affected product, problem and actions to be taken. Beckman Coulter will schedule a site visit to verify the reporting units of measurement for the laboratory within 90 days of this recall initiation. For questions call 800-854-3633.
Initiated
2018-03-02
Root cause
Under Investigation by firm
Terminated
2023-07-27
Distribution
Worldwide Distribution - US Nationwide and the countries of Brazil, Canada, Malaysia, Singapore, Taiwan, Thailand, Turkey
Recall number
Z-1643-2018
Recall status
Terminated
CDRH record ID
162677
FDA record created
2018-05-04
Recall event number
79575

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗