- FDA identifier
- Z-1644-2008
- Company in FDA record
- Emageon
- FDA classification date
- May 28, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Miscalculation: The calculation of the standard uptake value does not include the required calibration factor/decay correction factor. The resulting value is incorrectly calculated, which may lead to incorrect interpretation of tissue density. The calculation error exists in both the software and user guide.
Source facts
- Action
- Consignees were notified by Urgent Device Correction letters that were sent on 1/9/08 and 1/18/08. The letters instructed users not to use the standard uptake value in EVMS for diagnostic interpretation when viewing PET studies unless they received service bulletin 445, 446 or 7.0.1. Users were also instructed to replace pages in the user's guide with the pages provided and to affix the enclosed warning sticker on all workstations. Users will be contacted by the recalling firm to schedule a software patch.
- Initiated
- 2008-01-01
- Root cause
- Device Design
- Terminated
- 2009-06-11
- Distribution
- Nationwide Distribution including states of: MI, WI, CA, MN, AZ, CO, MN, MD, DC, MO, TX, IL, GA, FL, WI, SC, PA, NY, KS, TN, AL and OK.
- Recall number
- Z-1644-2008
- Recall status
- Terminated
- CDRH record ID
- 69669
- FDA classified
- 2008-05-28
- Recall event number
- 47526