Skip to content

Public FDA record

Z-1644-2014: Polyflux Revaclear Max, Capillary Dialyzer, Model Number 110634; Gambro Renal Products, Inc. Intended

Recall record Z-1644-2014 for Gambro Renal Products, Incorporated. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1644-2014
Company in FDA record
Gambro Renal Products, Incorporated
FDA classification date
May 21, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The firm received several complaints for internal blood leaks involving various lot numbers produced in late 2013.

Source facts

Action
The recall was initiated by an Urgent: Medical Device Recall letter delivered by UPS Overnight Delivery on April 16, 2014. The letter identified the affected product and reason for the recall. The firm is asking customers to remove and quarantine any remaining affected product from their inventory for collection and replacement. Customers are to also complete and return the Customer Reply Form to the Gambro Regulatory Affairs Department. Customers with affected product are to contact Gambro Customer Support or their Gambro representative to return the product. Questions should be directed to Customer Support at 1-800-651-2623.
Initiated
2014-04-16
Root cause
Storage
Terminated
2015-12-08
Distribution
Worldwide Distribution -- USA, including the territory of Puerto Rico, and the countries of Bermuda and Australia.
Recall number
Z-1644-2014
Recall status
Terminated
CDRH record ID
126757
FDA classified
2014-05-21
Recall event number
67996

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗