Skip to content

Public FDA record

Z-1645-2014: 550 TxT Treatment Table of the Digital Linear Accelerator (LINAC) To deliver X-ray radiation for therape

Recall record Z-1645-2014 for Siemens Medical Solutions USA, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1645-2014
Company in FDA record
Siemens Medical Solutions USA, Inc
FDA classification date
May 21, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Table may lose calibration during patient treatment, which may cause the possibility of mistreatment.

Source facts

Action
Siemens sent an Medical Device Correction, Field Safety Notice of Improvement to End users in March 2014, informing them that a Siemens representative would install a software update. By installation of this software update the issues were solved by the following actions: 1. A plausibility check was introduced for comparing the two independent encoder systems of each table axis. 2. The time between two encoder comparisons was reduced to 50ms and an error code was introduced. Please include this Field Safety Notice in your Digital Linear Accelerator System Owner Manual chapter "Safety Advisory Letters" where it should remain. The relevant National Competent Authority has been informed of this update. We regret any inconvenience that this may cause, and we thank you in advance for your understanding. Further questions please call (610) 219-6300.
Initiated
2014-04-01
Root cause
Software design
Terminated
2014-12-11
Distribution
USA ( nationwide) Distribution.
Recall number
Z-1645-2014
Recall status
Terminated
CDRH record ID
126663
FDA classified
2014-05-21
Recall event number
67957

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.