The DC power to the IFOSS collision detector control panel at your facility is being adversely affected by the magnetic field. Our investigation shows the magnetic field is causing the IFOSS control panel to malfunction.
Source facts
Action
IMRIS sent Service bulletin T200627-B01 Rev dated March 14, 2018. - through email to all affected customers. The letter identified the affected product, the problem, and the actions required to be taken by the customer. For questions contact IMRIS Customer Support 866-475-0525.
Initiated
2018-04-04
Root cause
Device Design
Terminated
2020-07-22
Distribution
Worldwide Distribution - US Nationwide in the states of OH, FL, NH, TX, Sweden, Canada
Recall number
Z-1645-2018
Recall status
Terminated
CDRH record ID
163559
FDA record created
2018-05-04
Recall event number
79816