Z-1645-2023: EarlyVue VS30 Vitals monitor with Software Versions A.00.02 and A.00.01 Product Number: 863380

Recall record Z-1645-2023 for Philips North America Llc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1645-2023
Company in FDA record
Philips North America Llc
FDA classification date
May 26, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

EarlyVue VS30 Vital Signs Monitor and EarlyVue VS30 Vitals monitor Missing Calibration Alarm, a risk of an inaccurate CO2 measurement which may lead to a failure to recognize a change in patient condition.

Source facts

Action
Philips issued Urgent Field Safety Notice Letter on 3/3/23. Letter states reason for recall, health risk and action to take: 1. Identify the last calibration date of the device and review your hospital maintenance plan. Determine if your device has E01 End Tidal CO2 and has not been calibrated within one year of the production calibration. If needed adjust the maintenance plan and incorporate the calibration of your affected device. a. The device should not be used until the calibration or maintenance plan has been reviewed. b. The calibration date can be accessed via the following steps on the device: Settings > Admin > Diagnostics > Page 2 of Diagnostics > Maintenance > 129 > CO2 Test. The image below shows the example of the screen to see the last calibration date of your device: 2. Review this URGENT Field Safety Notice in its entirety and pass this notice to all those who need to be aware within your organization or to any organizations where the potentially affected devices have been transferred. (If appropriate). 3. Complete and return attached form to Philips promptly to confirm receipt of the URGENT Field Safety Notice Letter, understanding of the issue, and required actions to be taken. 4. Actions planned by Philips Hospital Patient Monitoring to correct the problem A Philips representative will reach out to you to arrange a software upgrade to your monitor(s). Philips will start to schedule the upgrade upon formal release of the field action, which is planned for a six month implementation timeline. If you need any further information or support concerning this issue, please contact your local Philips representative
Initiated
2023-03-06
Root cause
Software design
Distribution
Nationwide Foreign: Belgium Brazil Canada Denmark Germany Guam Hong Kong Indonesia Italy Norway Poland Singapore Sweden Switzerland United Kingdom
Recall number
Z-1645-2023
Recall status
Open, Classified
CDRH record ID
199549
FDA classified
2023-05-26
Recall event number
92063

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗