Skip to content

Public FDA record

Z-1645-2026: BD Alaris" System with Guardrails" Suite MX with Point of Care Unit

Recall record Z-1645-2026 for CareFusion 303, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1645-2026
Company in FDA record
CareFusion 303, Inc.
FDA classification date
Mar 26, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to product labeling not indicating which Transport Layer Security (TLS) version that must be enabled on hospital network to ensure secure Wi-Fi communication with Point-of-Care Units

Source facts

Action
On 02/11/2026, the firm sent via US Postal mail, an "URGENT MEDICAL DEVICE PRODUCT CORRECTION" letter to customers informing them that labeling does not indicate which Transport Layer Security (TLS) versions must be enabled on the hospital network to ensure secure Wi-Fi communication with the Alaris and BD Alaris Point-of-Care Unit (PCU) when using enterprise-level Wi-Fi security. Without clear labeling, hospitals may unknowingly disable TLS 1.0, resulting in network connectivity failure and memory leaks which could lead to potential System Error. Customers/clinicians are instructed to: 1. Ensure the PCU is connected to the hospital network by viewing the wireless indicator icon on the front of the PCU. If the icon is illuminated or blinking, the PCU is connected. If it is not illuminated, the PCU is not connected. 2. If icon is not illuminated, follow established hospital protocols for wireless network downtime. 3. If a network connectivity failure persists, System Error may occur on the PCU. This error is indicated by a high priority alarm and red blinking arrow above the SYSTEM ON button. The customer must power down the PCU, then power on again. Any infusion that was ongoing at the time of System Error must be reprogrammed. 4. If using enterprise-level network security, customers/IT Security should be aware of which versions of TLS must be enabled to ensure connectivity. Refer to Appendix A for additional network administration guidance and compatibility. 5. Disseminate and post this notice within their facility to ensure that all concerned personnel, including IT and network connectivity owners or administrators, are notified and aware of this issue. For Questions/Assistance - contact the Recall Support Center at: Phone: 1-888-562-6018 Phone Hours: 7:00am to 4:00pm PT, Monday- Friday Email: SupportCenter@bd.com
Initiated
2026-02-11
Root cause
Under Investigation by firm
Distribution
Worldwide distribution: U.S.(Nationwide) to states of: AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MP. MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and WY O.U.S. (foreign) to countries of: Australia, Belgium, Bahrain, Botswana, Canada, China, Cuba, Germany, India, Israel, Italy, Japan, Korea, Kuwait, Malaysia, Mexico, Netherlands, New Zealand, Pakistan, Philippines, Qatar, Saudi Arabia, Singapore, South Africa, Switzerland, Taiwan, Turkey, United Arab Emirates, and United Kingdom.
Recall number
Z-1645-2026
Recall status
Open, Classified
CDRH record ID
218018
FDA classified
2026-03-26
Recall event number
98250

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗