Open original FDA record ↗Recall record
- FDA identifier
- Z-1646-2009
- Company in FDA record
- Disetronic Medical Systems, Inc.
- FDA classification date
- May 26, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The "up" and "down" buttons, which are used for changing the program in the menu or to administer additional insulin through a bolus delivery, may experience intermittent failure or completely cease to operate.
Source facts
- Action
- The firm issued a press release on 4/30/09 and sent letters with attached Urgent Medical Device Recall (Field Corrective Action) notification dated 5/4/09 to distributors, health care providers and users. Disetronic will replace pumps within the affected serial numbers (SN02119552 to SN10006093) that have intermittent or complete loss of function of the up and/or down button promptly upon contact from the customer with a new pump that is not affected by the potential button failure. For pumps within the affected serial numbers (SN02119552 to SN10006093) which are functioning properly, no replacement is available/possible at this time. Disetronic will gradually phase in replacement pumps for the remaining pumps within the affected range, as they become available. Customers are to complete the enclosed Customer Reply Card to acknowledge receipt of the notification. On 5/17/11, after being notified of the recall's ineffectiveness (in that only 40% of the pumps had been replaced), the firm re-issued recall notification letters including the aforementioned information to their consignees.
- Initiated
- 2009-04-30
- Root cause
- Device Design
- Terminated
- 2012-06-05
- Distribution
- Worldwide Distribution -- USA and Canada.
- Recall number
- Z-1646-2009
- Recall status
- Terminated
- CDRH record ID
- 81008
- FDA classified
- 2009-05-26
- Recall event number
- 51695