Skip to content

Public FDA record

Z-1647-2023: Philips Azurion System with software versions: R2.2.O, R2.2.1, R2.2.3, R2.2.5 and R2.2.6. Models/Model N

Recall record Z-1647-2023 for PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1647-2023
Company in FDA record
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
FDA classification date
Sep 30, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential Loss of X-ray-When the full disk capacity is reached, X-Ray functionality will cease to be available there will be a sudden interruption of the procedure

Source facts

Action
Philips issued Urgent Medical Device Letter to US consignees via e-mail on March 21, 2023 and certified email on March 29, 2023. Letters outside of the U.S.A. will be distributed through the local Philips Organizations. Letter states reason for recall, health risk and action to take: Keep this Medical Device Correction with the documentation of the system until Philips corrects your system. Circulate this notice to all users of the system so that they are aware of the issue. Return the attached reply form to Philips to confirm that the users of the system have reviewed and understood this Medical Device Correction. US: Please return this signed form via email to: IGT_Recalls@philips.com or alternatively via regular mail to Philips at 222 Jacobs Street, Cambridge, MA 02141 for the attention of Mr. Roland Telson / 3rd Floor Philips is working on a software release that will correct this issue (reference FCO72200528). In the interim, until this software is available and installed in your affected system(s), Philips will be removing the Log Trace Files from the affected systems to free up disk capacity (reference FCO72200529). Philips will be prioritizing these activities based on the time the affected software release has been installed in the Philips Azurion system. You will be contacted by your local Philips representative to schedule these activities. If you need any further information or support concerning this issue, contact your local Philips representative.
Initiated
2023-03-21
Root cause
Under Investigation by firm
Distribution
Nationwide Foreign: Algeria Argentina Armenia Australia Austria Bahrain Bangladesh Belarus Belgium Bolivia Brazil Bulgaria Cambodia Chile China Colombia Costa Rica C¿te D'Ivoire Croatia Cyprus Czech Republic Denmark Dominican Rep Ecuador Egypt Estonia Finland France French Polynesia Germany Greece Hong Kong Hungary India Indonesia Iran Iraq Ireland Israel Italy Japan Jordan Kazakhstan Kenya Korea, Republic of Kosovo Kuwait Latvia Lebanon Lithuania Macao Macedonia Malaysia Mexico Morocco Myanmar Nepal Netherlands New Zealand Nicaragua Nigeria Norway Oman Pakistan Palestine Palestine, State of Panama Peru Philippines Poland Portugal Puerto Rico Qatar Romania Russian Federation Saudi Arabia Senegal Serbia Singapore Slovakia Slovenia South Africa South Korea Spain Sri Lanka Sweden Switzerland Taiwan Thailand Tunisia Turkey Turkmenistan Ukraine United Kingdom Utd.Arab.Emir. Uzbekistan Vietnam Yemen
Recall number
Z-1647-2023
Recall status
Open, Classified
CDRH record ID
199808
FDA classified
2024-09-30
Recall event number
92165

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.