Potential Loss of X-ray-When the full disk capacity is reached, X-Ray functionality will cease to be available
there will be a sudden interruption of the procedure
Source facts
Action
Philips issued Urgent Medical Device Letter to US consignees via e-mail on March 21, 2023 and certified email on March 29, 2023. Letters outside of the U.S.A. will be distributed through the local Philips Organizations. Letter states reason for recall, health risk and action to take:
Keep this Medical Device Correction with the documentation of the system until Philips corrects your system.
Circulate this notice to all users of the system so that they are aware of the issue. Return the attached reply form to Philips to confirm that the users of the system have reviewed and understood this Medical Device Correction. US: Please return this signed form via email to: IGT_Recalls@philips.com or alternatively via regular mail to Philips at 222 Jacobs Street, Cambridge, MA 02141 for the attention of Mr. Roland Telson / 3rd Floor
Philips is working on a software release that will correct this issue (reference FCO72200528). In the interim, until this software is available and installed in your affected system(s), Philips will be removing the Log Trace Files from the affected systems to free up disk capacity (reference FCO72200529).
Philips will be prioritizing these activities based on the time the affected software release has been installed in the Philips Azurion system. You will be contacted by your local Philips representative to schedule these activities.
If you need any further information or support concerning this issue, contact your local Philips representative.
Initiated
2023-03-21
Root cause
Under Investigation by firm
Distribution
Nationwide
Foreign:
Algeria
Argentina
Armenia
Australia
Austria
Bahrain
Bangladesh
Belarus
Belgium
Bolivia
Brazil
Bulgaria
Cambodia
Chile
China
Colombia
Costa Rica
C¿te D'Ivoire
Croatia
Cyprus
Czech Republic
Denmark
Dominican Rep
Ecuador
Egypt
Estonia
Finland
France
French Polynesia
Germany
Greece
Hong Kong
Hungary
India
Indonesia
Iran
Iraq
Ireland
Israel
Italy
Japan
Jordan
Kazakhstan
Kenya
Korea, Republic of
Kosovo
Kuwait
Latvia
Lebanon
Lithuania
Macao
Macedonia
Malaysia
Mexico
Morocco
Myanmar
Nepal
Netherlands
New Zealand
Nicaragua
Nigeria
Norway
Oman
Pakistan
Palestine
Palestine, State of
Panama
Peru
Philippines
Poland
Portugal
Puerto Rico
Qatar
Romania
Russian Federation
Saudi Arabia
Senegal
Serbia
Singapore
Slovakia
Slovenia
South Africa
South Korea
Spain
Sri Lanka
Sweden
Switzerland
Taiwan
Thailand
Tunisia
Turkey
Turkmenistan
Ukraine
United Kingdom
Utd.Arab.Emir.
Uzbekistan
Vietnam
Yemen
Recall number
Z-1647-2023
Recall status
Open, Classified
CDRH record ID
199808
FDA classified
2024-09-30
Recall event number
92165
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.