An increase in complaints related to an inability to advance the guidewire through the lumen of the device for certain batches. There is evidence of the polyurethane layer delaminating and in some instances material detachment in the catheters inner lining.
Source facts
Action
Boston Scientific issued an Urgent Medical Device Removal notice to its consignees on 03/21/2024 via overnight delivery. The notice explained the issue, the potential risk, and requested the removal of the affected products. Distributors were directed to notify their consignees.
Initiated
2024-03-21
Root cause
Under Investigation by firm
Distribution
Worldwide distribution. US, US Territories, EMEA, Canada, LATAM
and Asia Pacific countries/regions.
Recall number
Z-1647-2024
Recall status
Open, Classified
CDRH record ID
206636
FDA classified
2024-05-10
Recall event number
94240