Open original FDA record ↗Recall record
- FDA identifier
- Z-1648-2011
- Company in FDA record
- Bio-Rad Laboratories Inc
- FDA classification date
- Mar 14, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The firm received customer complaints of increased positivity rates associated with the Anti-dsDNA test kits.
Source facts
- Action
- The firm, Bio-Rad, sent an "URGENT: MEDICAL DEVICE RECALL NOTIFICATION" letter dated March 2, 2010 to all customers. The letter described the product, problem and action to be taken by the customers. The customers were instructed to examine their inventory and quarantine the affected kit lots; discard all affected product in accordance with all local, state and federal disposal procedures; complete and return the enclosed MEDICAL DEVICE RECALL RESPONSE FORM via fax to: (510) 741-3954, attn: Bio-Rad CSD Regulatory Affairs Department, and to please re-evaluate all positive patient samples and retest if necessary. If you have any questions, contact Bio-Rad CSD Regulatory Affairs Department at 510-724-7000.
- Initiated
- 2010-03-02
- Root cause
- Nonconforming Material/Component
- Terminated
- 2011-03-15
- Distribution
- Worldwide distribution: USA and countries including: Canada, Carribean, Chile, France, Germany, Hong Kong, India, Italy, Japan, New Zealand, Panama, Singapore, South Korea and United Kingdom.
- Recall number
- Z-1648-2011
- Recall status
- Terminated
- CDRH record ID
- 92481
- FDA classified
- 2011-03-14
- Recall event number
- 56044