Open original FDA record ↗Recall record
- FDA identifier
- Z-1648-2012
- Company in FDA record
- Captiva Spine, Inc
- FDA classification date
- May 23, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Captiva Spine is recalling the SmartLOX Cervical Plate components 12 & 14 mm Drill Bits, Single Barrel Variable Drill Guides, and Double Barrel Fixed Drill Guides. The components are part of the instrumentation kit used to complete an anterior screw fixation of the cervical spine. There is a potential risk for the drills to not properly fit through the drill guides due to an overlap in the dime
Source facts
- Action
- Captiva Spine notified customers of the recall by letter on November 23, 2011, and advised that the affected products would be replaced. Customers were asked to return the drill guides and unused drills with the provided return labels. For questions customers should call 877-772-5571. For questions regarding this recall call 561-277-9480.
- Initiated
- 2011-11-22
- Root cause
- Nonconforming Material/Component
- Terminated
- 2013-11-18
- Distribution
- Nationwide Distribution including TX, IN, UT, AZ, GA, CA, AL, MI, and MN.
- Recall number
- Z-1648-2012
- Recall status
- Terminated
- CDRH record ID
- 109147
- FDA classified
- 2012-05-23
- Recall event number
- 61811