Skip to content

Public FDA record

Z-1648-2012: Accessories for the Captiva Spine SmartLOX Cervical Plate System. Products are labeled in part:

Recall record Z-1648-2012 for Captiva Spine, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1648-2012
Company in FDA record
Captiva Spine, Inc
FDA classification date
May 23, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Captiva Spine is recalling the SmartLOX Cervical Plate components 12 & 14 mm Drill Bits, Single Barrel Variable Drill Guides, and Double Barrel Fixed Drill Guides. The components are part of the instrumentation kit used to complete an anterior screw fixation of the cervical spine. There is a potential risk for the drills to not properly fit through the drill guides due to an overlap in the dime

Source facts

Action
Captiva Spine notified customers of the recall by letter on November 23, 2011, and advised that the affected products would be replaced. Customers were asked to return the drill guides and unused drills with the provided return labels. For questions customers should call 877-772-5571. For questions regarding this recall call 561-277-9480.
Initiated
2011-11-22
Root cause
Nonconforming Material/Component
Terminated
2013-11-18
Distribution
Nationwide Distribution including TX, IN, UT, AZ, GA, CA, AL, MI, and MN.
Recall number
Z-1648-2012
Recall status
Terminated
CDRH record ID
109147
FDA classified
2012-05-23
Recall event number
61811

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗