Skip to content

Public FDA record

Z-1648-2023: Cytrellis ellacor System with Micro-Coring Technology-Dermal Micro-Coring System System intended for the

Recall record Z-1648-2023 for Cytrellis Biosystems, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1648-2023
Company in FDA record
Cytrellis Biosystems, Inc.
FDA classification date
May 26, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Irregular core patterns could cause potential overlap of cores at the 8% setting, potentially leading to prolonged healing, irregularities in patient skin texture post-treatment, and/or scarring.

Source facts

Action
Cytrellis issued Medical Device Update / Product Notification Letter via email on 4/7/23. Letter states reason for recall, heath risk and action to take: Important information is included in the new Quick Reference Guide, with emphasis on the following points: "DO NOT impede tubing on handpiece. "DO NOT apply forceful downward pressure. "DO NOT TOUCH needle cartridge tubing while treating patient. Your PDM will be reaching out soon to set up time with you and your staff to review the Quick Reference Guide, and we would also like to invite you to join our medical team in one of two live informative webinars to review best practices and answer any questions that you might have. These webinars will be hosted by our Medical Director, Dr. Ashish Bhatia, along with the Cytrellis Medical Affairs team. Invitations to the webinars will be sent shortly. For additional information or inquiries, please email medicalinformation@cytrellis.com
Initiated
2023-04-07
Root cause
Use error
Distribution
Nationwide
Recall number
Z-1648-2023
Recall status
Completed
CDRH record ID
199841
FDA classified
2023-05-26
Recall event number
92187

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗