Open original FDA record ↗Recall record
- FDA identifier
- Z-1648-2023
- Company in FDA record
- Cytrellis Biosystems, Inc.
- FDA classification date
- May 26, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Irregular core patterns could cause potential overlap of cores at the 8% setting, potentially leading to prolonged healing, irregularities in patient skin texture post-treatment, and/or scarring.
Source facts
- Action
- Cytrellis issued Medical Device Update / Product Notification Letter via email on 4/7/23. Letter states reason for recall, heath risk and action to take: Important information is included in the new Quick Reference Guide, with emphasis on the following points: "DO NOT impede tubing on handpiece. "DO NOT apply forceful downward pressure. "DO NOT TOUCH needle cartridge tubing while treating patient. Your PDM will be reaching out soon to set up time with you and your staff to review the Quick Reference Guide, and we would also like to invite you to join our medical team in one of two live informative webinars to review best practices and answer any questions that you might have. These webinars will be hosted by our Medical Director, Dr. Ashish Bhatia, along with the Cytrellis Medical Affairs team. Invitations to the webinars will be sent shortly. For additional information or inquiries, please email medicalinformation@cytrellis.com
- Initiated
- 2023-04-07
- Root cause
- Use error
- Distribution
- Nationwide
- Recall number
- Z-1648-2023
- Recall status
- Completed
- CDRH record ID
- 199841
- FDA classified
- 2023-05-26
- Recall event number
- 92187