Z-1649-2023: Trilogy Evo, Garbin Evo, and Aeris EVO, Model Numbers DS2110X11B, IN2110X15B, JP2110X16B, AU2110X15B, LA2

Recall record Z-1649-2023 for Philips Respironics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1649-2023
Company in FDA record
Philips Respironics, Inc.
FDA classification date
Jun 2, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Environmental debris may accumulate on the internal machine flow sensor causing partial occlusion which may impact accurate delivery of pressure, volume, or flow. A malfunction caused by this issue can result in patient harm up to hypoxemia if not addressed by the care provider.

Source facts

Action
An URGENT Medical Device Correction notification letter was sent to customers on 5/1/23. Please take the actions below in order to prevent risk for your patients. Actions that must be taken by the user in order to prevent risks for patients. To help prevent accumulation of debris on the machine flow sensor: Use the Philips approved particulate filter which prevents a significant majority of airborne aerosols and particulate from entering the device. This filter must be replaced between patients and monthly as indicated in the Instructions for Use. Using this filter was previously optional. This is now required. Use the air inlet filter as indicated in the Instructions for Use. Installation of the particulate filter will not require a change to therapy settings. To help detect change in therapy: Set appropriate alarms based on ventilation mode, such as Low Tidal Volume, Low Minute Ventilation, Low Inspiratory Pressure, and High Inspiratory Pressure. Check Proximal Pressure and External Flow Sensor Failed alarms can also alert the user to this issue. These are non settable alarms. The device will issue a low priority Inlet Filter(s) Blocked alarm if therapy is reduced due to filter occlusion. The device will continue to function if this occurs, but the user must be aware to rinse the air inlet filter and replace the particulate filter if this occurs. Observe instructions for any alarm, especially Ventilator Service Required or Ventilator Inoperative. If the situation cannot be resolved, use alternative ventilation equipment. Ensure ventilator dependent patients have access to alternative ventilation equipment, such as a back up ventilator or manual resuscitator. Distribute this notice to all employees in your organization that need to be aware. Train device users within your network to use the air inlet and particulate filters on all patients. Send this notice to any organization an impacted device was sold or distributed to. If you need any further informatio
Initiated
2023-05-01
Root cause
Under Investigation by firm
Distribution
US Nationwide. Global Distribution.
Recall number
Z-1649-2023
Recall status
Open, Classified
CDRH record ID
199470
FDA classified
2023-06-02
Recall event number
92048

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗