Skip to content

Public FDA record

Z-1649-2026: Philips Respironics Trilogy Evo Universal, Software Version 1.05.15.00. Continuous home-use ventilator de

Recall record Z-1649-2026 for Philips Respironics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1649-2026
Company in FDA record
Philips Respironics, Inc.
FDA classification date
Apr 2, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

In some situations, the Obstruction Alarm does not trigger within the timeframe dictated by the relevant standards and may take up to four breaths.

Source facts

Action
An IMPORTANT PRODUCT NOTICE was mailed to consignees 6/5/24 about obstruction alarm compliance failure in Trilogy Evo devices. Refer to the most current version of the IFU in the event of an obstruction. A software update to correct this issue is pending. An UPDATED: URGENT FIELD SAFETY NOTICE was mailed 3/2/26 with recall information for three failure modes possible in all Trilogy Evo, Trilogy EVO O2, Trilogy Evo Universal, and Trilogy EV300. Using non-pneumatic nebulizers with Trilogy Evo Platform ventilators may result in a discrepancy between the set tidal volume and the tidal volume received by the patient. This can lead to under-delivery of therapy to the patient. Update all devices with software version 1.05.15.00 through the My Philips for Professionals and InCenter websites. Stop use of non-pneumatic nebulizers, including vibrating mesh nebulizers, with all affected devices. Distributors give your customers the provided recall notice and response form. DME and Homecare Customers call 1-800-345-6443 (option 2, option 1) for assistance. Hospital Customers call 1-800-722-9377 (option 2) for assistance. **An UPDATED: URGENT FIELD SAFETY NOTICE was mailed to consignees 4/24/26 with recall information for three failure modes possible in all Trilogy Evo, Trilogy EVO O2, Trilogy Evo Universal, and Trilogy EV300 devices. It clarifies and revises guidance about the use of non-pneumatic nebulizers with Trilogy Evo Platform ventilators communicated in the 3/2/26 update. This does not change this previously communicated information: Nebulized aerosol deposition on the internal flow sensor (FSN 2024-CC-SRC-013 - Flow Sensor Nebulized Aerosol Deposition), and Obstruction Alarm timing (IPN 2024-CC-SRC-002 - Obstruction Alarm). Consignees review the latest version of the User Manual Addendum, give the safety notice to all who interact with affected devices, and forward the notice to those to whom product was further distributed.
Initiated
2026-03-02
Root cause
Software design
Distribution
Worldwide Distribution. US Nationwide, Argentina, Austria, Australia, Bosnia and Herzegovina, Bangladesh, Belgium, Bulgaria, Bolivia, Brazil, Bahamas, Canada, Switzerland, Chile, China, Colombia, Czech Republic, Germany, Denmark, Algeria, Ecuador, Estonia, Egypt, Spain, Finland, France, Georgia, French Guiana, Gibraltar, Greece, Guatemala, Hong Kong, Croatia, Hungary, Indonesia, Ireland, Israel, India, Italy, Jersey, Jordan, Japan, Kenya, South Korea, Kuwait, Lebanon, Lithuania, Morocco, Montenegro, North Macedonia, Myanmar, Martinique, Malta, Mexico, Malaysia, Nigeria, Netherlands, Norway, Nepal, New Zealand, Oman, Panama, Peru, French Polynesia, Philippines, Pakistan, Poland, Portugal, Paraguay, Qatar, Reunion, Romania, Serbia, Russia, Saudi Arabia, Sweden, Singapore, Slovenia, Slovakia, Thailand, Tunisia, Turkey, Taiwan, Ukraine, United Arab Emirates, United Kingdom, Vietnam, Mayotte, South Africa, Zimbabwe.
Recall number
Z-1649-2026
Recall status
Open, Classified
CDRH record ID
219028
FDA classified
2026-04-02
Recall event number
98540

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.