Skip to content

Public FDA record

Z-1650-2012: Celsus prosthetic foot, Product Number 103658-00. The Celsus is intended for low impact level lower limb

Recall record Z-1650-2012 for College Park Industries, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1650-2012
Company in FDA record
College Park Industries, Inc.
FDA classification date
May 23, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

An incorrect shorter part ( pyramid) was installed in the foot. A foot of this size and configuration requires a longer pyramid. Customers requiring the extra firm configuration may have a minimal risk of component failure, which could be implicated in a potential fall.

Source facts

Action
All customers who received affected product were initially contacted via phone on April 5, 2012 and made aware of the recall, including instructions to notify customers immediately and arrange replacements. A "Recall Notice" letter dated April 24, 2012 was sent to customers describing the problem and the recommended actions. Contact reference person is Kimberly Light, 1 800-728-7950, info@college-park.com.
Initiated
2012-04-05
Root cause
Mixed-up of materials/components
Terminated
2012-07-31
Distribution
Worldwide Distribution - USA, including the states of CA, MA, ME, WA, TX, VA, AZ and the country of England.
Recall number
Z-1650-2012
Recall status
Terminated
CDRH record ID
108756
FDA classified
2012-05-23
Recall event number
61701

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗