Open original FDA record ↗Recall record
- FDA identifier
- Z-1650-2012
- Company in FDA record
- College Park Industries, Inc.
- FDA classification date
- May 23, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
An incorrect shorter part ( pyramid) was installed in the foot. A foot of this size and configuration requires a longer pyramid. Customers requiring the extra firm configuration may have a minimal risk of component failure, which could be implicated in a potential fall.
Source facts
- Action
- All customers who received affected product were initially contacted via phone on April 5, 2012 and made aware of the recall, including instructions to notify customers immediately and arrange replacements. A "Recall Notice" letter dated April 24, 2012 was sent to customers describing the problem and the recommended actions. Contact reference person is Kimberly Light, 1 800-728-7950, info@college-park.com.
- Initiated
- 2012-04-05
- Root cause
- Mixed-up of materials/components
- Terminated
- 2012-07-31
- Distribution
- Worldwide Distribution - USA, including the states of CA, MA, ME, WA, TX, VA, AZ and the country of England.
- Recall number
- Z-1650-2012
- Recall status
- Terminated
- CDRH record ID
- 108756
- FDA classified
- 2012-05-23
- Recall event number
- 61701