- FDA identifier
- Z-1650-2016
- Company in FDA record
- Neusoft Medical Systems Co., Ltd.
- FDA record creation date
- May 11, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
It was found by R&D in April, 2014 that there was a defect in NeuViz 64 system software version 1.0.5+P09. The icons of side decubitus didn't meet the patient position description, when the scanner position was set to "Right" in the system setting - scanner options.
Source facts
- Action
- An upgraded software version (1.0.5.2619+P11) was released on May 7, 2015. The FCO was released to the service engineers of Neusoft Medical Systems USA, which is subsidiary of the reporting firm, and the engineers performed FCO for each affected device together with the distributors' engineers. The corrections (FCOs to upgrade software) on all the affected systems in the USA have been completed. At present (4/27/16) there is no longer this issue on the systems, so no further actions are needed for customers/distributors. However, the recalling firm shall inform the direct accounts of this fix before May 5, 2016. The recalling firm will submit the communication evidence ASAP once they have evidence of the communication. For further questions, please call (281) 453-1205.
- Initiated
- 2015-05-02
- Root cause
- Device Design
- Terminated
- 2016-11-04
- Distribution
- US Distribution - Including Puerto Rico and the states of IL, CT, SC, NE
- Recall number
- Z-1650-2016
- Recall status
- Terminated
- CDRH record ID
- 145070
- FDA record created
- 2016-05-11
- Recall event number
- 73832