Elekta sent an "IMPORTANT FIELD SAFETY NOTICE" dated April 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter instructs customers to carry out the recommended weekly physics check of the X-ray and light field size calibration. If an error is found, customers are instructed to contact their local Elekta service engineer.
Initiated
2012-04-19
Root cause
Nonconforming Material/Component
Terminated
2014-10-06
Distribution
Nationwide Distribution including Puerto Rico.
Recall number
Z-1651-2012
Recall status
Terminated
CDRH record ID
109148
FDA classified
2012-05-23
Recall event number
61812