Open original FDA record ↗Recall record
- FDA identifier
- Z-1651-2014
- Company in FDA record
- Heraeus Kulzer, LLC.
- FDA classification date
- May 23, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The firm discovered that the Gluma Desensitizer Power Gel can be inadvertently extruded from the application syringe all at once when it is supposed to extrude on the desired area a small amount at a time. If extruded all at once, product could reach the dental patients mucosa, face or eyes or the dental professional could be exposed. This would cause irritation, burning or blistering.
Source facts
- Action
- Heraeus Kulzer sent a Voluntary Recall letter dated April 22, 2014 to all affected doctors and supply centers. The letter identified the affected product, problem and actions to be taken. The letter instructed customers to: 1. Inspect your inventory, both in storage and in use, and identify the lot number on your kits and syringes. Refer to attached images. 2. If you identify any Gluma Desensitizer PowerGel with the aforementioned kit or syringe lot numbers, immediately cease usage of this product and call Technical Services at 1 (574) 299-5409 3. Please advise if you forwarded this product to any other end users. We will need their contact information inClusive of name, address and phone number. For questions regarding this notification, please call our Technical Service team at (574) 299-5409.
- Initiated
- 2014-04-23
- Root cause
- Nonconforming Material/Component
- Terminated
- 2014-11-05
- Distribution
- US Nationwide Distribution in the states of GA, TX, IL, CA, WA, WI, HI, MO, FL, NJ, TN, PA, NY, NC, OH, IN, CT, KY, AZ, NV, IA,
- Recall number
- Z-1651-2014
- Recall status
- Terminated
- CDRH record ID
- 127168
- FDA classified
- 2014-05-23
- Recall event number
- 68223