Skip to content

Public FDA record

Z-1651-2014: Gluma Desensitizer Power Gel Product Usage: Usage: This is a class 2 medical device. For reduction or e

Recall record Z-1651-2014 for Heraeus Kulzer, LLC.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1651-2014
Company in FDA record
Heraeus Kulzer, LLC.
FDA classification date
May 23, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The firm discovered that the Gluma Desensitizer Power Gel can be inadvertently extruded from the application syringe all at once when it is supposed to extrude on the desired area a small amount at a time. If extruded all at once, product could reach the dental patients mucosa, face or eyes or the dental professional could be exposed. This would cause irritation, burning or blistering.

Source facts

Action
Heraeus Kulzer sent a Voluntary Recall letter dated April 22, 2014 to all affected doctors and supply centers. The letter identified the affected product, problem and actions to be taken. The letter instructed customers to: 1. Inspect your inventory, both in storage and in use, and identify the lot number on your kits and syringes. Refer to attached images. 2. If you identify any Gluma Desensitizer PowerGel with the aforementioned kit or syringe lot numbers, immediately cease usage of this product and call Technical Services at 1 (574) 299-5409 3. Please advise if you forwarded this product to any other end users. We will need their contact information inClusive of name, address and phone number. For questions regarding this notification, please call our Technical Service team at (574) 299-5409.
Initiated
2014-04-23
Root cause
Nonconforming Material/Component
Terminated
2014-11-05
Distribution
US Nationwide Distribution in the states of GA, TX, IL, CA, WA, WI, HI, MO, FL, NJ, TN, PA, NY, NC, OH, IN, CT, KY, AZ, NV, IA,
Recall number
Z-1651-2014
Recall status
Terminated
CDRH record ID
127168
FDA classified
2014-05-23
Recall event number
68223

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗