Skip to content

Public FDA record

Z-1651-2025: stryker Latitude EV HUMERAL REVISION STEM TRIAL, L: 150mm, Size: LARGE, Side: RIGHT, REF 9030103; Prosth

Recall record Z-1651-2025 for Tornier, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1651-2025
Company in FDA record
Tornier, Inc
FDA classification date
Apr 22, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A specific lot of Latitude Humeral Trial Stems has the incorrect direct part marking. " Laser Marked: 9030103, Humeral Stem Trial 150mm Large / Right, Yellow color indicator (incorrect) " Device Actually: 9030092, Humeral Stem Trial 150mm Medium / Left, Red color indicator (correct)

Source facts

Action
Stryker issued an "URGENT MEDICAL DEVICE RECALL" notice to its consignees on 03/03/2025 via Trackwise (email). The notice explained the issue, potential risk, and requested the following: "1. Immediately check your internal inventory to locate the product listed on the attached business reply form, remove them from their point of use, and isolate/quarantine the unit(s) to prevent accidental use. 2. Return the enclosed business reply form by email to confirm receipt of this notification/document product segregation. a. Response is required, even if you may not have any physical inventory on site anymore. Completing this form will allow us to update our records and will also negate the need for us to send any further unnecessary communications on this matter. Therefore, please complete even if you no longer have any of the subject devices in your physical inventory. 3. Upon receipt of the completed business reply form, Stryker will contact you to arrange for the return of your product(s). 4. Maintain awareness of this communication internally until all required actions have been completed within your facility. 5. If you have further distributed the affected product, please notify the applicable parties at once about this recall. a. If possible, inform us if any of the subject devices have been distributed to other organizations, including contact details so that we can inform the recipients appropriately. b. If you are a distributor, note that you are responsible for notifying your affected customers. 6. Please inform us of any adverse event and/or report them to the Health/Competent Authorities in accordance with current regulations. For questions or concerns, please contact memphis.fieldaction@stryker.com."
Initiated
2025-03-03
Root cause
Under Investigation by firm
Distribution
OUS (International) Distribution to countries: Sweden, Australia, United Kingdom, Colombia, Germany
Recall number
Z-1651-2025
Recall status
Open, Classified
CDRH record ID
213028
FDA classified
2025-04-22
Recall event number
96494

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗