Open original FDA record ↗Recall record
- FDA identifier
- Z-1652-2008
- Company in FDA record
- Spacelabs Healthcare, Incorporated
- FDA classification date
- Sep 18, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Telemetry Modules are failing to alarm for low heart rate and asystole.
Source facts
- Action
- On 07/18/08 Spacelabs Healthcare issued a "URGENT-MEDICAL DEVICE CORRECTION" letter, dated July 16, 2008, to all customers who received the Spacelabs Medical Telemetry Receiver-Model 90478. The letter states circumstances in which the alarm fails, how to test each module for high rate alarm prior to patient use and a temporary work around if the alarm failure exists.` Letters have a receipt acknowledgement to ensure each customer receives the letter. Spacelabs Technical Support will personally install the new software, version V3.05.07EN, in all Spacelabs Medical Telemetry Receivers Model 90478. Technical Support will also be responsible for responding to customer technical questions and ensuring each customer is upgraded regarding this medical device recall. For assistance contact SpaceLabs Medical at 1-425-657-7200.
- Initiated
- 2008-07-18
- Root cause
- Software design
- Terminated
- 2011-03-21
- Distribution
- Nationwide and Internationally
- Recall number
- Z-1652-2008
- Recall status
- Terminated
- CDRH record ID
- 72451
- FDA classified
- 2008-09-18
- Recall event number
- 48985