Siemens Radiation Oncology became aware that customers may be using the Siemens Linear Accelerator in combination with stereotactic accessories which have not been validated as being compatible with Siemens LINAC models.
Source facts
Action
Siemen sent an Urgent Medical Device Correction letter to all affected customers. The letter identified the affected product, problem and actions to be taken. For questions contact your local Elekta representative.
Initiated
2014-04-01
Root cause
Labeling design
Terminated
2015-03-16
Distribution
US Nationwide Distribution in the states of AL, ME, PA, IL, and CA.
Recall number
Z-1652-2014
Recall status
Terminated
CDRH record ID
126662
FDA classified
2014-05-23
Recall event number
67956