Z-1652-2022: FGFR RGQ RT-PCR Kit (RUO, not IVD) REF 8747010

Recall record Z-1652-2022 for Qiagen Sciences LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1652-2022
Company in FDA record
Qiagen Sciences LLC
FDA record creation date
Aug 29, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Increased occurrence rate of false positive results, lead to a false positive sample result that could subsequently be incorrectly reported from the laboratory. A false positive result could lead to the risk of the patient being exposed to an inappropriate or suboptimal anti-cancer treatment and/or unnecessarily experiencing side effects

Source facts

Action
Qiagen issued July 20,2022 letter ( REF 874721) via email (A) US customers: notify customers by Urgent Medical Device Correction Letter; (B) CE customers: notify customers by Urgent Field Safety Notice (C) RUO (not IVD) customers: notify customers by Important Notice. Letter states reason for recall, health risk and action to take: For the continued use of REF 874721 therascreen FGFR RGQ RT-PCR Kit LOT 172017274 , please note the following criteria: Samples with the Overall sample result No Alteration Detected should be regarded as correct and no further actions are needed. Samples with the FGFR Alteration Detected results with Individual target result S249C Mutation Detected can be regarded as correct. This also includes samples with additional alterations detected. All other FGFR Alteration Detected results aside from S249C Mutation Detected should be disregarded and a retest for confirmatory positive result should be performed using the extracted RNA if available, or re-extracted if it is not. The following results should be assigned to the affected samples after the retest: l Retest result is positive for the same target(s)/ alteration(s): regard the result for the corresponding sample as positive. l Retest result is positive for different target(s)/ alteration(s) in the retest compared to the initial result: regard the result as indeterminate. l Retest is negative: the result for the corresponding sample should be regarded as negative l Retested sample is within an invalid run: repeat the run. l Result for individual sample is invalid in the retest: regard the result as indeterminate Forward this information to all individuals and departments within your organization who are using REF 874721 therascreen FGFR RGQ RT-PCR Kit. l If you are not the end user, please forward this notice to the product end user. l Review this notice with your laboratory/medical director. l Complete the Acknowledgement of Receipt Form attached to this letter by 28 July
Initiated
2022-07-20
Root cause
Under Investigation by firm
Distribution
IN, NC, MN, TX Foreign: Belgium, China, Germany, Jordan, Italy, Netherlands, Malaysia, Spain, Thailand
Recall number
Z-1652-2022
Recall status
Open, Classified
CDRH record ID
195434
FDA record created
2022-08-29
Recall event number
90691

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗