Open original FDA record ↗Recall record
- FDA identifier
- Z-1654-2008
- Company in FDA record
- Radiometer America Inc
- FDA record creation date
- Jun 25, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Reported calibration errors are not displayed on the analyzers parameter bar screen or on the patient results when the calibration number is between 32768 - 65535 and 98303 - 131070. This may cause the operator to perform measurements and obtain questionable results.
Source facts
- Action
- On February 28, 2008, a Field Action Notes Letter was issued advising them about a Mandatory Software Update. Customers were notified of the problem by fax and letter (Fed Ex). The Recall Letter instructed the user to disable the parameter bar feature and only rely on the traffic light system indicator which properly indicates system errors. The letter includes instructions to temporarily fix the problem and notice of a update of the software module that updates the parameter bar. If you have any questions contact Vincent Sigmund at (440)-871-8900 ext. 209.
- Initiated
- 2008-01-18
- Root cause
- Software design
- Terminated
- 2012-02-24
- Distribution
- Worldwide Distribution - USA and Canada - Products were sold in all fifty (50) US States. including twenty-nine 29 government accounts and 138 Canadian accounts.
- Recall number
- Z-1654-2008
- Recall status
- Terminated
- CDRH record ID
- 67928
- FDA record created
- 2015-06-25
- Recall event number
- 46686