Open original FDA record ↗Recall record
- FDA identifier
- Z-1655-2010
- Company in FDA record
- Computerized Medical Systems Inc
- FDA classification date
- May 26, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Under certain circumstances, XiO will over-estimate dose in the region where the backup jaw is inside the MLC opening
Source facts
- Action
- Elekta issued an undated "Important Safety Notice" to all affected sites via e-mail on 4/20/2010. The notice was also sent to the sites on 4/6/10 and 4/12/10 (for 4 sites) before the firm made the decision to recall. The letter explains the correct functioning of the system, the problem, when and why it occurs, the clinical impact, and the work around. Although not stated in the Safety Notice, a self-addressed postcard was enclosed for the customer to return as evidence they received and understood the notice. Sites that do not return the card will be contacted by phone.
- Initiated
- 2010-04-15
- Root cause
- Software design
- Terminated
- 2012-06-07
- Distribution
- Worldwide Distribution: USA, Canada, Australia, Spain, Italy, the Netherlands, United Kingdom, Taiwan, Malaysia, France, Ireland, and Israel.
- Recall number
- Z-1655-2010
- Recall status
- Terminated
- CDRH record ID
- 90674
- FDA classified
- 2010-05-26
- Recall event number
- 55361